DS-1050 Series Image Spectrophotometer

DS-1050 Series Image Spectrophotometer


Specifications
Details

Objective evaluation of "unmeasurable materials"

The DigiSpec® series image spectrophotometer is a groundbreaking color measurement instrument that can be used in virtually all areas of color measurement. The image spectrophotometer can acquire color and spectral data for each point in the entire image per minute, with an area as small as 0.0016mm² (0.04mm*0.04mm), between square inches.
The image spectrophotometer uses an LED or xenon light source as the illumination source, which ensures that the traceability data is consistent with the previous generation of conventional spectrophotometers.

Complex color and mixed color measurement
The minimum measurement is 0.0016mm ²
Maximum repeatability deab ≤ 0.01
Traceability data consistency
 

Epoch making color measuring instrument

DigiSpec® Series image spectrophotometer can be applied to almost all color measurement fields

Textile industry: yarn, printing, lace, zipper, embroidery, etc

Construction industry: rock color paint, real stone paint, etc

Printing industry: wood grain paper, wallpaper, etc

Restoration of Cultural Relics: murals, calligraphy and painting, ceramics, etc

Others: buttons, jewelry, etc

 

Every inch is worth a penny

DigiSpec ® The color and spectral data of each point in the whole image can be obtained. In the area of 0.04mm * 0.04mm, the minimum measurement can be 0.0016mm ²

DigiSpec® It is helpful for the consistency of color evaluation in the whole production process

 

Digital transmission, fast and accurate color reproduction

The maximum value of measurement repeatability deab ≤ 0.01, whether measuring white standard plate or black standard plate digiscpec & reg; It can achieve the maximum measurement repeatability, deab ≤ 0.01 ※, and put forward a new standard of benchmark significance for desktop spectrophotometer

The perfect calibration structure and algorithm provide completely reliable short-term measurement stability and long-term measurement stability

 

※The surface color of different materials is greatly affected by temperature. Deab ≤ 0.01 is an extremely precise measurement stability condition. In the inspection of digiscpec & reg; When measuring repeatability, be sure to ensure the surface temperature stability of the measured object

 

Traceability data consistency

Using LED or xenon light source as lighting source can ensure the consistency of traceability data with the previous generation of traditional spectrophotometer

There are 20 + standard light sources and 30 + measurement indicators. SCI / SCE measurement conditions cover all the functions of the previous generation of spectrophotometer, and can provide the measurement parameters and conditions of the traditional spectrophotometer
 

Model DS-1000 DS-1050 DS-1100
Geometry Dual beam D / 8 ° , SCI / SCE 
Light source Full spectrum LED High precision like sunlight full spectrum LED High precision analog fluorescent xenon lamp light source
Sphere diameter 152mm/6in
Wavelength range

400nm-700nm

400nm-1000nm

Reporting interval 10nm 2.5nm
Photometric range 0-200%, resolution 0.01%

Repeated measurement of

30 times of double flash
White color difference (CIE LAB)※

ΔE*ab≤0.03(Maximum) ΔE*ab≤0.01(Maximum)

Inter-Instrument consistency for

reflectance measurement
 (CIELAB)※※

0.4 0.25
Aperture plate Lav (30mm lighting, 25mm measurement), the user can set the measurement area and customize the caliber of any size
Test conditions

Compliance with standards:CIE No.15、GB/T 3978、GB 2893GB/T 18833、ISO7724/1、DIN5033 Teil7、

JIS Z8722 condition C、ASTM E1164、ASTM-D1003-07

Sensor CMOS array sensor

Silicon based metal oxide image sensor

Spectroscopic modes grating Ultra high precision holographic transmission volume grating
Image resolution 300dpi 500dpi(high resolution) 600dpi(Ultra high resolution)
Minimum measurement area better than 0.01mm² (0.1*0.1mm) better than 0.004mm² (0.06*0.06mm) better than 0.0016mm² (0.04*0.04mm)
Observer 2 ° standard observer and 10 ° standard observer
 light source A,C,D50,D55,D65,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,U30,DLF,NBF,TL83,TL84
Color space L*a*b,L*C*h,Hunter Lab,Yxy,XYZ
Color index

WI(ASTM E313-00,ASTM E313-73,CIE/ISO,AATCC,Hunter,Taube,Berger Stensby),

YI(ASTM D1925,ASTM E313-00,ASTM E313-73),Tint(ASTM E313-00),Isochromatic index milm, color fastness,

Color fastness, ISO brightness, R457, a density, t density, e density, m density, hiding power, strength

Colour-difference formula ΔE*ab,ΔE*CH,ΔE*uv,ΔE*cmc,ΔE*94,ΔE*00,ΔEab(Hunter),555 hue classification
Test time Less than 8 seconds Less than 5 seconds
Working temperature and humidity 5 ~ 40 ° C, no condensation below 80% relative humidity (at 35 ° C)
Storage temperature and humidity -20 ~ 45 ° C, no condensation below 80% relative humidity (at 35 ° C)
Accessories Power adapter, data cable, whiteboard
Interface USB 3.0

※ The measuring caliber is 25 * 25MM. After the instrument is calibrated, measure the BCRA white correction plate 30 times at an interval of 10 seconds

※※ Average value of measurement results of 12 BCRA ceramic standard color plates

 

 

 



 

View more about DS-1050 Series Image Spectrophotometer on main site